Therapeutic Ultrasound

Introduction

 

Ultrasound has been a part of clinical practice since sometime back in the 1950's, and remains a popular and evidenced intervention for a range of clinical problems. Shah and Farrow (2012) provide an insight into its current clinical popularity as does the widely cited paper by Pope et al (1995). General (textbook) reviews and explanations can be found in Watson and Young (2008) and Robertson et al (2006) amongst others.

 

There are myriad therapy ultrasound machines available, from the small, portable devices, through to the multimodal machines which include ultrasound as one of the available options, examples are illustrated below.

​

Introduction
US Energy

Ultrasound (US) is a form of MECHANICAL energy, not electrical energy and therefore strictly speaking, not really electrotherapy at all but does fall into the Electro Physical Agents grouping. Mechanical vibration at increasing frequencies is known as sound energy. The normal human sound range is from 16Hz to something approaching 15-20,000 Hz (in children and young adults). Beyond this upper limit, the mechanical vibration is known as ULTRASOUND. The frequencies used in therapy are typically between 1.0 and 3.0 MHz (1MHz = 1 million cycles per second).

Sound waves are LONGITUDINAL waves consisting of areas of COMPRESSION and RAREFACTION. Particles of a material, when exposed to a sound wave will oscillate about a fixed point rather than move with the wave itself. As the energy within the sound wave is passed to the material, it will cause oscillation of the particles of that material. Clearly any increase in the molecular vibration in the tissue can result in heat generation, and ultrasound  can be used to produce thermal changes in the tissues, though current usage in therapy does not focus on this phenomenon (Williams 1987, Baker et al 2001, ter Haar 1999, Nussbaum 1997, Watson 2000, 2008, 2020). In addition to thermal changes, the vibration of the tissues appears to have effects which are generally considered to be non thermal in nature, though, as with other modalities (e.g. Pulsed Shortwave) there must be a thermal component however small. As the US wave passes through a material (the tissues), the energy levels within the wave will diminish as energy is transferred to the material. The energy absorption and attenuation characteristics of US waves have been documented for different tissues (see absorption section).

US Waves
Ultrasound Waves :

 

FREQUENCY -     the number of times a particle experiences a complete compression/rarefaction cycle in 1 second. Typically 1 or 3 MHz (though there are devices which operate in the kHz range - see comments on Low Frequency / Longwave Ultrasound at the end of this page).

 

WAVELENGTH - the distance between two equivalent points on the waveform in the particular medium. In an ‘average tissue’ the wavelength @ 1MHz would be 1.5mm and @ 3 MHz would be 0.5 mm.

 

VELOCITY - the velocity at which the wave (disturbance) travels through the medium. In a saline solution, the velocity of US is approximately 1500 m sec-1 compared with approximately 350 m sec-1 in air (sound waves can travel more rapidly in a more dense medium). The velocity of US in most tissues is thought to be similar to that in saline.

 

These three factors are related, but are not constant for all types of tissue. Average figures are most commonly used to represent the passage of US in the tissues. Typical US frequencies from therapeutic equipment are 1 and 3 MHz though some machines produce additional frequencies (e.g. 0.75 and 1.5 MHz) and the ‘Longwave’ ultrasound devices operate at several 10’s of kHz (typically 40-50,000Hz – a much lower frequency than ‘traditional US’ but still beyond human hearing range.

 

The mathematical representation of the relationship is V = F.l

where V = velocity, F = frequency and l is the wavelength.

Fields and BNR
Ultrasound Beam, Near Field, Far Field and Beam Non Uniformity

 

The US beam is not uniform and changes in its nature with distance from the transducer. The US beam nearest the treatment head is called the NEAR field, the INTERFERENCE field or the Frenzel zone. The behaviour of the US in this field is far from regular, with areas of significant interference. The US energy in parts of this field can be many times greater than the output set on the machine (possibly as much as 12 to 15 times greater). The size (length) of the near field can be calculated using r2/l where r= the radius of the transducer crystal and l = the US wavelength according to the frequency being used (0.5mm for 3MHz and 1.5mm for 1.0 MHz).

 

As an example, a 'crystal' with a diameter of 25mm operating at 1 MHz will have a near field/far field boundary at : Boundary = 12.5mm2/1.5mm » 10cm  thus the near field (with greatest interference) extends for approximately 10 cm from the treatment head when using a large treatment head and 1 MHz US. When using higher frequency US, the boundary distance is even greater. Beyond this boundary lies the Far Field or the Fraunhofer zone. The US beam in this field is more uniform and gently divergent. The ‘hot spots’ noted in the near field are not significant. For the purposes of therapeutic applications, the far field is effectively out of reach.

 

One quality indicator for US applicators (transducers) is a value attributed to the Beam Nonuniformity Ratio (BNR). This gives an indication of this near field interference. It describes numerically the ratio of the intensity peaks to the mean intensity. For most applicators, the BNR would be approximately 4 - 6 (i.e. that the peak intensity will be 4 or 6 times greater than the mean intensity). It is considered inappropriate to use a device with a BNR value of 8.0 or more. Because of the nature of US, the theoretical best value for the BNR is thought to be around 4.0 though some manufacturers claim to have overcome this limit and effectively reduced the BNR of their generators to 1.0.

 

Some recent papers (Alvarenga et al 2010; Gutierrez et al 2010; Straub et al, 2008 and Johns et al 2007) have considered some of the inaccuracies associated with current machines and Pye (1996) presents some worrying data with regards the calibration of machines in clinical use in the UK.

Transmission
Ultrasound Transmission through the Tissues

 

All materials (tissues) will present an impedance to the passage of sound waves. The specific impedance of a tissue will be determined by its density and elasticity. In order for the maximal transmission of energy from one medium to another, the impedance of the two media needs to be as similar as possible. Clearly in the case of US passing from the generator to the tissues and then through the different tissue types, this can not actually be achieved. The greater the difference in impedance at a boundary, the greater the reflection that will occur, and therefore, the smaller the amount of energy that will be transferred. Examples of impedance values can be found in the literature e.g. Robertson et al 2007, Ward 1986.

 

The difference in impedance is greatest for the steel/air interface which is the first one that the US has to overcome in order to reach the tissues. To minimise this difference, a suitable coupling medium has to be utilised. If even a small air gap exists between the transducer and the skin the proportion of US that will be reflected approaches 99.998% which means that there will be no effective transmission.

​

The coupling media used in this context include water, various oils, creams and gels. Ideally, the coupling medium should be fluid so as to fill all available spaces, relatively viscous so that it stays in place, have an impedance appropriate to the media it connects, and should allow transmission of US with minimal absorption, attenuation or disturbance. For a good discussion regarding coupling media, see Casarotto et al 2004, Docker et al 1982, Griffin, 1980, Klucinec et al 2000 and Williams 1987. At the present time the gel based media are preferable to the oils and creams. Water is an effective media and can be used as an alternative but clearly it fails to meet the above criteria in terms of its viscosity. There is no realistic (clinical) difference between the gels in common clinical use (Poltawski and Watson 2007). The addition of active agents (e.g. anti-inflammatory drugs) to the gel is widely practiced, but remains incompletely researched. We are currently evaluating this intervention further.

Bacterial Contamination of Ultrasound Treatment Heads and Gel Sources

As a matter of (clinical) interest, the US treatment should be cleaned with an alcohol based swab (not just wiped with tissue) between treatments (Schabrun et al, 2006) to minimise the potential transmission of microbial agents between patients. Spratt et al (2014) sampled ultrasound treatment heads and gel bottles (in the USA) reporting over 50% of gel bottles with contamination, some of which were positive for MRSA. Some 35% of the ultrasound treatment heads tested also demonstrated contamination, though none with MRSA. The authors report that employment of adequate disinfection techniques was effective in significantly reducing these levels (they used Protex, Parker Laboratories).

Ultrasound Application - The Critical Angle

In addition to the reflection that occurs at a boundary due to differences in impedance, there will also be some refraction if the wave does not strike the boundary surface at 90°.

 

Essentially, the direction of the US beam through the second medium will not be the same as its path through the original medium - its pathway is angled. The critical angle for US at the skin interface appears to be about 15°. If the treatment head is at an angle of 15° or more to the plane of the skin surface, the majority of the US beam will travel through the dermal tissues (i.e. parallel to the skin surface) rather than penetrate the tissues as would be expected.

Absorption
Ultrasound Absorption and Attenuation :

 

The absorption of US energy follows an exponential pattern - i.e. more energy is absorbed in the superficial tissues than in the deep tissues. In order for energy to have an effect it must be absorbed, and at some point this must be considered in relation to the US dosages applied to achieve certain effects (ter Haar, 1999, Watson, 2008, 2020, Watson and Young, 2008).

 

Because the absorption (penetration) is exponential, there is (in theory) no point at which all the energy has been absorbed, but there is certainly a point at which the US energy levels are not sufficient to produce a therapeutic effect. As the US beam penetrates further into the tissues, a greater proportion of the energy will have been absorbed and therefore there is less energy available to achieve therapeutic effects. The half value depth is often quoted in relation to US and it represents the depth in the tissues at which half the surface energy is available. These will be different for each tissue and also for different US frequencies. The table below gives some indication of typical (or average) half value depths for therapeutic ultrasound. (after Hoogland 1995)

As it is difficult, if not impossible to know the thickness of each of these layers in an individual patient, average half value depths are employed for each frequency

3MHz = 2cm

1MHz = 4cm

These values (after Low & Reed) are not universally accepted (see Ward 1986) and some research (as yet unpublished) suggests that in the clinical environment, they may be significantly lower.

​

To achieve a particular US intensity at depth, account must be taken of the proportion of energy which has been absorbed by the tissues in the more superficial layers. The table gives an approximate reduction in energy levels with typical tissues at two commonly used frequencies, and more detailed information is found in the dose calculation material

​

As the penetration (or transmission) of US is not the same in each tissue type, it is clear that some tissues are capable of greater absorption of US than others. Generally, the tissues with the higher protein content will absorb US to a greater extent, thus tissues with high water content and low protein content absorb little of the US energy (e.g. blood and fat) whilst those with a lower water content and a higher protein content will absorb US far more efficiently. Tissues can be ranked according to their relative tissue absorption and this is critical in terms of clinical decision making (Watson, 2008 2020).

As the penetration (or transmission) of US is not the same in each tissue type, it is clear that some tissues are capable of greater absorption of US than others. Generally, the tissues with the higher protein content will absorb US to a greater extent, thus tissues with high water content and low protein content absorb little of the US energy (e.g. blood and fat) whilst those with a lower water content and a higher protein content will absorb US far more efficiently. Tissues can be ranked according to their relative tissue absorption and this is critical in terms of clinical decision making (Watson, 2008 2020).

Although cartilage and bone are at the upper end of this scale, the problems associated with wave reflection mean that the majority of US energy striking the surface of either of these tissues is likely to be reflected. The best absorbing tissues in terms of clinical practice are those with high collagen content – LIGAMENT, TENDON, FASCIA, JOINT CAPSULE, SCAR TISSUE (Watson 2000, 2008, Watson & Young, 2008, ter Haar 1999, Nussbaum 1998, Frizzel & Dunn 1982)

The application of therapeutic US to tissues with a low energy absorption capacity is less likely to be effective than the application of the energy into a more highly absorbing material. Recent evidence of the ineffectiveness of such an intervention can be found in Wilkin et al (2004) and Markert et al (2005) whilst application in tissue that is a better absorber will, as expected, result in a more effective intervention (e.g. Sparrow et al 2005, Leung et al 2004).

The physiological effects of ultrasound are almost identical to those of Pulsed Shortwave and Laser therapy – the key difference however, is that ultrasound energy is preferentially absorbed in different tissue to the other modalities – as summarised in the diagram.

Pulsing
Pulsed Ultrasound

Most machines offer the facility for pulsed US output, and for many clinicians, this is a preferable mode of treatment. Until recently, the pulse duration (the time during which the machine is on) was almost exclusively 2ms (2 thousandths of a second) with a variable off period. Some machines now offer a variable on time though whether this is of clinical significance has yet to be determined. Typical pulse ratios are 1:1 and 1:4 though others are available (see dose calculations). In 1:1 mode, the machine offers an output for 2ms followed by 2ms rest. In 1:4 mode, the 2ms output is followed by an 8ms rest period. The adjacent diagram illustrates the effect of varying the pulse ratio.

The effects of pulsed US are well documented and this type of output is preferable especially in the treatment of the more acute lesions. Some machines offer pulse parameters that do not appear to be supported from the literature (e.g. 1:9; 1:20).

Some manufacturers describe their pulsing in terms of a percentage rather than a ratio (1:1 = 50%  1:4 = 20% etc). An equivalence table is provided for convenience. The proportion of time that the machine is ON compared with OFF is a relevant factor in dosage calculations and further details are included in the dose calculation support material

Pulse Frequency

A point of confusion amongst many therapists is the 'frequency' facility offered on some ultrasound machines. The pulse ratio (duty cycle) is at say 1:4 (20%), but there is an option to alter the pulse frequency (i.e. how many ultrasound pulses are delivered per second). This is achieved by adjusting the DURATION of the pulses. Typically, these are at 2ms, thus on a 1:4 ratio, the machine is ON for 2ms and then OFF for 8ms. It takes 10ms to complete one 'cycle' (ON : OFF), and thus 100 such cycles are completed in a second, so the machine claims to deliver ultrasound at 100Hz. If the pulse duration is increased from 2ms to say 4ms, then on a 1:4 ratio, the machine will be ON for 4ms followed by 16ms OFF, thus taking 20ms to complete a cycle and hence only 50 such cycles being delivered per second and the setting on the machine will be for ultrasound pulsing at 50Hz. There is no evidence that I can find to suggest that one mode of operation has any clinical advantage over another. The 2ms pulse time is 'normal' and is encountered on most machines (which effectively means that 100 pulses of ultrasound energy will be delivered per second.

Thermal
Therapeutic Ultrasound Thermal and Non Thermal Effects Overview

 

One of the therapeutic effects for which ultrasound has been used is in relation to tissue healing. It is suggested that the application of US to injured tissues will, amongst other things, speed the rate of healing & enhance the quality of the repair (Watson 2006). The following information is intended to provide a summary of some of the essential research in this field together with some possible mechanisms through which US treatments may achieve these changes. It is not intended to be a complete explanation of these phenomena or a comprehensive review of the current literature.  It may, none the less, provide some useful basic information for clinical application. Some of the 'wider' applications for ultrasound therapy (e.g. drug delivery, chemotherapy potentiation) are usefully reviewed in Paliwal and Mitragotri (2008).

 

The therapeutic effects of US are generally divided into: THERMAL & NON‑THERMAL.

Thermal:

In thermal mode, US will be most effective in heating the dense collagenous tissues and will require a relatively high intensity, preferably in continuous mode to achieve this effect.

 

Many papers have concentrated on the thermal effectiveness of ultrasound, and much as it can be used effectively in this way when an appropriate dose is selected (continuous mode >0.5 W cm-2), the focus of this paper will be on the non thermal effects. Both Nussbaum (1998) and ter Haar (1999) have provided some useful review material with regards the thermal effects of ultrasound. Comparative studies on the thermal effects of ultrasound have been reported by several authors (e.g. Draper et al 1993, 1995a,b, Leonard et al 2004) with some interesting, and potentially useful results. Further work continues in our research centre with a comparison of contact heating and longwave ultrasound (Meakins and Watson, 2006) and comparison of different US regimes combined with US (Aldridge and Watson – in preparation)

 

It is too simplistic to assume that with a particular treatment application there will either be thermal or non thermal effects. It is almost inevitable that both will occur, but it is furthermore reasonable to argue that the dominant effect will be influenced by treatment parameters, especially the mode of application i.e. pulsed or continuous. Baker et al (2001) have argued the scientific basis for this issue coherently.

 

Lehmann (1982) suggests that the desirable effects of therapeutic heat can be produced by US. It can be used to selectively raise the temperature of particular tissues due to its mode of action.  Among the more effectively heated tissues are periosteum, collagenous tissues (ligament, tendon & fascia) & fibrotic muscle (Dyson 1981).  If the temperature of the damaged tissues is raised to 40‑45°C, then a hyperaemia will result, the effect of which will be therapeutic. In addition, temperatures in this range are also thought to help in initiating the resolution of chronic inflammatory states (Dyson & Suckling 1978). Most authorities currently attribute a greater importance to the non-thermal effects of US as a result of several investigative trials in the last 15 years or so.

Non-Thermal:

 

The non-thermal effects of US are now attributed primarily to a combination of CAVITATION and  ACOUSTIC STREAMING (ter Haar 1999, 2008 Baker et al 2001, Williams 1987, Watson 2020). There appears to be little by way of convincing evidence to support the notion of MICROMASSAGE though it does sound rather appealing.

 

CAVITATION in its simplest sense relates to the formation of gas filled voids within the tissues & body fluids. There are 2 types of cavitation ‑ STABLE & UNSTABLE which have very different effects. STABLE CAVITATION does seem to occur at therapeutic doses of US.  This is the formation & growth of gas bubbles by accumulation of dissolved gas in the medium.  They take apx. 1000 cycles to reach their maximum size. The `cavity' acts to enhance the acoustic streaming phenomena (see below) & as such would appear to be beneficial. UNSTABLE (TRANSIENT) CAVITATION is the formation of bubbles at the low pressure part of the US cycle. These bubbles then collapse very quickly releasing a large amount of energy which is detrimental to tissue viability. There is no evidence at present to suggest that this phenomenon occurs at therapeutic levels if a good technique is used. There are applications of US that deliberately employ the unstable cavitation effect (High Intensity Focussed Ultrasound or HIFU) but it is beyond the remit of this summary.

ACOUSTIC STREAMING is described as a small scale eddying of fluids near a vibrating structure such as cell membranes & the surface of stable cavitation gas bubble (Dyson & Suckling 1978). This phenomenon is known to affect diffusion rates & membrane permeability. Sodium ion permeability is altered resulting in changes in the cell membrane potential. Calcium ion transport is modified which in turn leads to an alteration in the enzyme control mechanisms of various metabolic processes, especially concerning protein synthesis & cellular secretions.

​

The result of the combined effects of stable cavitation and acoustic streaming is that the cell membrane becomes ‘excited’ (up regulates), thus increasing the activity levels of the whole cell. The US energy acts as a trigger for this process, but it is the increased cellular activity which is in effect responsible for the therapeutic benefits of the modality (Watson 2000, 2008, Dinno et al 1989, Leung et al 2004, El Ghamrawy et al 2019)

MICROMASSAGE is a mechanical effect which appears to have been attributed less importance in recent years. In essence, the sound wave travelling through the medium is claimed to cause molecules to vibrate, possibly enhancing tissue fluid interchange & affecting tissue mobility. There is little, if any hard evidence for this often cited principle

US Repair
Ultrasound Application in relation to Tissue Repair

 

The process of tissue repair is a complex series of cascaded, chemically mediated events that lead to the production of scar tissue that constitutes an effective material to restore the continuity of the damaged tissue. The process is more complex than be described here, but there are several interesting recent papers and reviews including (Wener & Grose 2003, Toumi & Best 2003, Watson 2003, 2006, 2020, Hill et al 2003, Neidlinger-Wilke et al 2002, Lorena et al 2002, Latey 2001, Velnar et al 2009, Hauser et al 2013).

 

The various phases of tissue repair can be usefully represented by the ‘blocks’ in the figure below. The division into Bleeding, Inflammatory, Proliferative and Remodelling phases is almost arbitrary in that from a tissue perspective, this is in fact one continuous series of events, with a change in emphasis with time. Further details, reviews and reference materials can be found in the publications identified above or from the TISSUE REPAIR pages on this site.

INFLAMMATION:

During the inflammatory phase, US has a stimulating effect on the mast cells, platelets, white cells with phagocytic roles and the macrophages (Nussbaum 1997, ter Haar 1999, Fyfe & Cahal 1982, Maxwell 1992, Watson 2008, 2020; Li et al 2003). For example, the application of ultrasound induces the degranulation of mast cells, causing the release of arachidonic acid which itself is a precursor for the synthesis of prostaglandins and leukotreine – which act as inflammatory mediators (Mortimer & Dyson 1988, Nussbaum 1997, Leung et al 2004). By increasing the activity of these cells, the overall influence of therapeutic US is certainly pro-inflammatory rather than anti-inflammatory. The benefit of this mode of action is not to ‘increase’ the inflammatory response as such (though if applied with too greater intensity at this stage, it is a possible outcome (Ciccone et al 1991), but rather to act as an ‘inflammatory optimiser’ (Watson 2007, 2008, 2020).

 

The inflammatory response is essential to the effective repair of tissue, and the more efficiently the process can complete, the more effectively the tissue can progress to the next phase (proliferation). Studies which have tried to demonstrate the anti inflammatory effect of ultrasound have failed to do so (e.g.El Hag et al 1985 Hashish 1986, 1988), and have suggested that US is ineffective. It is effective at promoting the normality of the inflammatory events, and as such has a therapeutic value in promoting the overall repair events (ter Haar 99, Watson 2008, 2020, Bajpai et al 2018; Sahu et al 2019). A further benefit is that the inflammatory chemically mediated events are associated with stimulation of the next (proliferative) phase, and hence the promotion of the inflammatory phase also acts as a promoter of the proliferative phase.

Employed at an appropriate treatment dose, with optimal treatment parameters (intensity, pulsing and time), the benefit of US is to make as efficient as possible to earliest repair phase, and thus have a promotional effect on the whole healing cascade. For tissues in which there is an inflammatory reaction, but in which there is no ‘repair’ to be achieved, the benefit of ultrasound is to promote the normal resolution of the inflammatory events, and hence resolve the ‘problem’ This will of course be most effectively achieved in the tissues that preferentially absorb ultrasound – i.e. the dense collagenous tissues.

PROLFERATION:

During the proliferative phase (scar production) US also has a stimulative effect (cellular up regulation), though the primary active targets are now the fibroblasts, endothelial cells and myofibroblasts (Ramirez et al 1997, Mortimer and Dyson 1988, Young & Dyson 1990, Young & Dyson 1990b, Nussbaum 1997, 1998, Dyson & Smalley 1983, Maxwell 1992, Watson and Young 2008, Watson 2007, 2008, 2020, Ng 2011; de Oliveira Perrunchini et al 2020). These are all cells that are normally active during scar production and US is therefore pro-proliferative in the same way that it is pro-inflammatory – it does not change the normal proliferative phase, but maximises its efficiency – producing the required scar tissue in an optimal fashion. Harvey et al (1975) demonstrated that low dose pulsed ultrasound increases protein synthesis and several research groups have demonstrated enhanced fibroplasia and collagen synthesis (Enwemeka et al 1989, 1990, Turner et al 1989, Huys et al 1993, Ramirez et al 1997, Warden et al 2006, Zhang et al 2004). Recent work has identified the critical role of numerous growth factors in relation to tissue repair, and some accumulating evidence has identified that therapeutic US has a positive role to play in this context (e.g. Leung et al, 2006; Lovric et al 2013; Li et al, 2002; McBrier et al 2007; Reher et al 1999; Tsai et al 2006) and also with heat shock proteins (Nussbaum and Locke 2007). There are over 350 papers which identify the influence of US on the production of collagen and associated proliferative events.

REMODELLING:

During the remodelling phase of repair, the somewhat generic scar that is produced in the initial stages is refined such that it adopts functional characteristics of the tissue that it is repairing. A scar in ligament will not ‘become’ ligament, but will behave more like a ligamentous tissue. This is achieved by a number of processes, but mainly related to the orientation of the collagen fibres in the developing scar (Culav et al 1999, Gomez et al 1991, Watson, 2003) and also to the change in collagen type, from predominantly Type III collagen to a more dominant Type I collagen. The remodelling process is certainly not a short duration phase – research has shown that it can last for a year or more – yet it is an essential component of quality repair (El Batouty et al 1986, ter Haar 1987)

 

The application of therapeutic ultrasound can influence the remodelling of the scar tissue in that it appears to be capable of enhancing the appropriate orientation of the newly formed collagen fibres and also to the collagen profile change from mainly Type III to a more dominant Type I construction, thus increasing tensile strength and enhancing scar mobility (Nussbaum 1998, Wang 1998, de Oliveira Perrunchini et al 2020, Watson 2020). Ultrasound applied to tissues enhances the functional capacity of the scar tissues (Nussbaum 1998, Huys et al 1993, Tsai et al 2006, 2011, Yeung et al 2006). The role of ultrasound in this phase may also have the capacity to influence collagen fibre orientation as demonstrated in an elegant study by Byl et al (1996), though their conclusions were quite reasonably somewhat tentative.

The application of ultrasound during the inflammatory, proliferative and repair phases is not of value because it changes the normal sequence of events, but because it has the capacity to stimulate or enhance these normal events and thus increase the efficiency of the repair phases (ter Haar 99, Watson 2007, 2008, Watson & Young, 2008). It would appear that if a tissue is repairing in a compromised or inhibited fashion, the application of therapeutic ultrasound at an appropriate dose will enhance this activity. If the tissue is healing ‘normally’, the application will, it would appear, speed the process and thus enable the tissue to reach its endpoint faster than would otherwise be the case. The effective application of ultrasound to achieve these aims is dose dependent.

Other Applications
Other Applications :

 

There are a growing number of ‘other applications’ for ultrasound energy ranging from tumour ablation – using High Intensity Focussed Ultrasound (or HIFU) though to stimulated related of encapsulated systemic drugs. HIFU is beyond the scope of this review which is mainly concerned with 'standard' therapy ultrasound.

 

LIPUS (Low Intensity Pulsed Ultrasound) for Fracture Repair – there is a wealth of research information in this area (summarised on the specific Ultrasound page on this site if you want the detail). The NICE guidelines (Jan 2013) are supportive of this clinical application. Essentially, the application of very low dose ultrasound over a fracture (whether healing normally or delayed or non union) can be of significant benefit. The main clinical issue is that the effective ‘dose’ is actually lower than most therapy machines can deliver – which is frustrating! Higher intensity ultrasound over a fracture can initiate a strong pain response – which is useful when it comes to using the modality to locate potential stress fractures (see below). A useful recent review (fully detailed on the web pages) can be found at Zura et al (2015). At the current time, there are over 1000 papers on the database which link the use of LIPUS with fracture healing and associated issues.

 

The LIPUS ultrasound field is starting to expand - working on the principle that if it is so effective for fracture healing, then it might also be effective for soft tissue lesions and other musculoskeletal problems. Much of this work is still in development, though results and publications are anticipated (reviewed in Khanna et al, 2009). Examples of additional bone related papers include El Bialy et al (2002) who evaluated its effects during distraction osteogenesis.

Low Frequency (Longwave) Ultrasound : There is a limited range of experimental, lab and clinical evidence with regards the specific use of ultrasound in the kHz range of frequencies (over the 20kHz hearing upper limit, but below the MHz frequencies more commonly employed in therapy. Ahmadi et al (2012) have usefully reviewed both bioeffects and safety issues at this frequency range. A separate Low Frequency (Longwave) Ultrasound information page and pdf download are available from the web pages (www.electrotherapy.org) for those interested in this field. Longwave (kHz) ultrasound is gaining an additional use in terms of wound debridement. There is a separate page which considers this application mode.

Therapy Ultrasound as a Diagnostic Tool for Stress Fractures : The use of therapy ultrasound machines as a means to 'diagnose' whether there is a stress fracture is a technique what has been employed sporadically for many years. Some claim (anecdotally) that it is very effective whilst others dismiss the technique. Essentially, a 'strong' ultrasound application (typically 1MHz; 1.0 W cm-2; continuous) is applied over the suspect area. If there is a stress fracture (or other significant bony injury) it is common for a sharp pain to be felt by the patient. It can be a useful technique when working without immediate access to XRay, MRI or other higher level imaging technology. The research reports covering this technique are mixed. The more recent papers include Papalada et al (2012) who suggest that the technique is reproducible and reliable whilst Khatri et al (2008) suggest that the technique is not as accurate as MRI - no argument - it is not claimed to be - but that does not make it invalid as a quick, cost effective and easily accessible assessment for the therapist. Other papers include Devereaux et al (1984) and Delacerda (1981).

Ultrasound Therapy for Wound Healing : There have been a range of research papers over the years which have set out to evaluate the benefits (or otherwise) of ultrasound therapy as a means to stimulate healing in chronic wounds (typically venous ulcers and pressure sores). Whilst some research has not demonstrated significant clinical benefit, others have clearly done so, and therefore, as with other wound based electrophysical agent applications, it is likely to be a dose dependent response. Cullum et al (2010) contributed a Cochrane review on this topic, though only 8 trials were included (I have over 400 papers the in one way or another consider ultrasound and wound healing). Of the 6 papers which employed MHz ultrasound, 5 demonstrated significant benefit, though as these trials employed small samples, it was suggested that the evidence was not sufficiently strong to constitute support for the therapy (in Cochrane terms). A quick search of the literature will identify numerous papers, which a swift review of the Cullum et al paper will give you rapid access to some useful leads as will the slightly older, but none the less useful Uhlemann et al (2003) publication which considers both MHz and kHz versions of the modality. Ennis et al (2005) provide a useful RCT based publication demonstrating beneficial outcomes. Low frequency (longwave) ultrasound (including MIST Therapy where the ultrasound is delivered through a saline 'mist') is being increasingly employed as an evidenced  wound debridement treatment (e.g. Bell et al, 2008; Stanisic et al, 2005) whilst Serena et al (2009) and Harris et al (2014) identify its potential benefit in relation to wound bacterial counts. Examples of recent, well constructed trials for ultrasonic treatment of pressure ulcers can be found in Polak et al (2014) and for diabetic related wounds in Bajpal et al 2018.

The use of US at trigger points has been used clinically for some time and is well supported from the anecdotal evidence. A study by Srbely et al (2007) raises some interesting points and demonstrates a measurable benefit. Other studies in this area include Sarrafzadeh et al (2012); Unalan et al (2011); Draper et al (2010); Aguilera et al (2009); Majlesi and Unalan (2004); Manca et al (2014). More recently Morishita et al (2014) have demonstrated some interesting effects of ultrasound (to trapezius), stretch and pain effects, which may link to trigger point applications.

 

A recent review and meta analysis (Zeng et al, 2014) evaluated continuous and pulsed ultrasound applications in knee osteoarthritis (OA) management. It was shown that pulsed ultrasound is more effective in terms of pain management and functional improvement compared with control conditions. The use of continuous ultrasound was effective only in terms of pain management.

 

There is an increasing body of evidence which supports the use of Ultrasound in chronic sinusitis and rhinosinusitis. This has been usefully reviewed in Bartley et al (2014) and detailed clinical trials can be found in Ansari et al (2014); Young et al (2010); Naghdi et al (2008); Ansari et al (2007). Doses applied in these trials were typically at 1MHz; 0.5-1.0W cm-2; pulsed mode; 15-18 minutes; 3 x weekly; 10 sessions with significant clinical benefit.

'STATUS' Ultrasound : Enraf (www.enraf-nonius.com) have recently introduced a 'new' method of delivering traditional ultrasound therapy which employs a stationary treatment applicator which is  held in place using a vacuum system. There is currently limited clinical research with regards any different effect - none are expected - the advantage being that (a) it is safe and (b) the therapist does not need to deliver the treatment in the classic way - they can affix the applicator and proceed with other jobs. The STATUS application employs 'traditional' ultrasound (as described in this paper) - it is sometimes confused with LIPUS (see above) given that the application is made with a stationary treatment applicator. This is not a LIPUS treatment.

 

Home based Ultrasound : there is an increasing availability of home based ultrasound treatment options. The (potential) advantage is that the therapist need not use clinic time to deliver the treatment, and secondly, if ultrasound is at its most effective when delivered daily, it becomes a realistic option. The disadvantage of most of these portable, handheld devices is that (currently) they rarely offer a choice of treatment dose, thus reducing their value. Some new systems are becoming available on which there are dose selection options, making them potentially useful (e.g. www.ultralieve.com). It is anticipated that is will be a treatment option which increases in popularity and value in the near future.

References
References

 

Aguilera, F. J. et al. (2009). "Immediate effect of ultrasound and ischemic compression techniques for the treatment of trapezius latent myofascial trigger points in healthy subjects: a randomized controlled study." J Manipulative Physiol Ther 32(7): 515-520.

 

Ahmadi, F. et al (2012). "Bio-effects and safety of low-intensity, low-frequency ultrasonic exposure." Prog Biophys Mol Biol 108(3): 119-138.

 

Alvarenga, A. V. and R. P. B. Costa-Félix (2010). "Metrological aspects on therapeutic ultrasound parameters: effective radiating area and non-uniformity ration." Phys Procedia 3(1): 643-649.

 

Anon (1983). "Ultrasonic assessment of stress fractures." BMJ 286: 1479.

 

Ansari, N. et al. (2007). "Physiotherapy for chronic rhinosinusitis: The use of continuous ultrasound." Int J Ther Rehabil 14(7): 306-310.

 

Ansari, N. et al. (2012). "A randomized, double-blind clinical trial comparing the effects of continuous and pulsed ultrasound in patients with chronic rhinosinusitis." Physiother Theory Pract 28(2): 85-94.

 

Bajpai, A. et al (2018) Effects of Non-thermal, Non-cavitational Ultrasound Exposure on Human Diabetic Ulcer Healing and Inflammatory Gene Expression in a Pilot Study Ultrasound Med Biol 44(9): 2043

 

Baker, K. G., et al. (2001). "A review of therapeutic ultrasound: biophysical effects." Phys Ther 81(7): 1351-8.

 

Bartley, J. et al. (2014). "Therapeutic ultrasound as a treatment modality for chronic rhinosinusitis." Current Infectious Disease Reports 16(3): 398.

 

Bell, A. L. and J. Cavorsi (2008). "Noncontact ultrasound therapy for adjunctive treatment of nonhealing wounds: retrospective analysis." Phys Ther 88(12): 1517-1524

 

Byl, N. N., et al. (1996). "Effects of ultrasound on the orientation of fibroblasts: an in-vitro study." Eur-J-Phys-Med-Rehabil 6(6): 180-4.

 

Casarotto, R. A., et al. (2004). "Coupling agents in therapeutic ultrasound: acoustic and thermal behavior." Archives of Physical Medicine and Rehabilitation 85(1): 162-5.

 

Ciccone, C. D., et al. (1991). "Effects of ultrasound and trolamine salicylate phonophoresis on delayed-onset muscle soreness." Phys Ther 71(9): 666-75; discussion 675-8.

 

Culav, E. M., et al. (1999). "Connective tissues : Matrix composition and its relevance to physical therapy." Physical Therapy 79(3): 308-319.

 

Cullum N et al. (2010) Therapeutic ultrasound for venous leg ulcers. Cochrane Database of Systematic Reviews  DOI: 10.1002/14651858.CD001180.pub3

 

de Oliveira Perrucini, P. D. et al (2020) Anti-Inflammatory and Healing Effects of Pulsed Ultrasound Therapy on Fibroblasts Am J Phys Med Rehabil 99(1): 19

 

Delacerda, F. G. (1981). "A case study: application of ultrasound to determine a stress fracture of the fibula." J Orthop Sports Phys Ther 2(3): 134-136.

 

Devereaux, M. D. et al. (1984). "The diagnosis of stress fractures in athletes." Jama 252(4): 531-533.

 

Dingemanse, R. et al. (2014). "Evidence for the effectiveness of electrophysical modalities for treatment of medial and lateral epicondylitis: a systematic review." British Journal of Sports Medicine 48(12): 957-965.

 

Dinno, M., et al. (1989). "The significance of membrane changes in the safe and effective use of therapeutic and diagnostic ultrasound." Phys Med Biol 34: 1543-1552.

 

Docker, M. F. et al. (1982). "Ultrasound couplants for physiotherapy." Physiotherapy 68(4): 124-125.

 

Draper, D. et al (1995). "Rate of temperature increase in human muscle during 1Mhz and 3Mhz continuous ultrasound." J Orthopaedic and Sports Physical Therapy 22(4): 142-150.

 

Draper, D. et al. (1993). "A comparison of temperature rise in human calf muscle following applications of underwater and topical gel ultrasound." JOSPT 17: 247-251.

 

Draper, D. O. and M. D. Ricard (1995). "Rate of temperature decay in human muscle following 3 MHz ultrasound: the stretching window revealed." Journal of Athletic Training 30(4): 304-7.

 

Draper, D. O. et al. (1995). "Temperature changes in deep muscles of humans during ice and ultrasound therapies: an in vivo study." Journal of Orthopaedic & Sports Physical Therapy 21(3): 153-7.

 

Draper, D. O. et al. (2010). "Thermal ultrasound decreases tissue stiffness of trigger points in upper trapezius muscles." Physiotherapy Theory and Practice 26(3): 167-172.

 

Dunn, F. and L. A. Frizzell (1982). Bioeffects of ultrasound. Therapeutic Heat and Cold. F. Lehmann J, Rehabilitation Medicine Library: 386-.

 

Dyson, M. and J. Niinikoski (1982). "Stimulation of tissue repair by therapeutic ultrasound." Infections in Surgery: 37-44.

 

Dyson, M. and D. Smalley (1983). Effects of ultrasound on wound contraction. Ultrasound Interactions in Biology & Medicine. R. Millner, E. Rosenfeld and U. Cobet. New York, Plenum Press: 151-158.

 

Dyson, M. and J. Suckling (1978). "Stimulation of tissue repair by ultrasound: A survey of the mechanisms involved." Physiotherapy 64(4): 105-108.

 

El-Batouty, M. F. et al. (1986). "Comparative evaluation of the effects of ultrasonic and ultraviolet irradiation on tissue regeneration." Scand J Rheumatology 15: 381-386.

 

El-Bialy, T. et al. (2002). "The effect of pulsed ultrasound on mandibular distraction." Ann Biomed Eng 30(10): 1251-1261.

 

El Ghamrawy, A. et al (2019) Acoustic Streaming in a Soft Tissue Microenvironment Ultrasound Med Biol 45(1): 208

 

El Hag, M. et al. (1985). "The anti-inflammatory effect of dexamethazone and therapeutic ultrasound in oral surgery." Br J Oral & Maxillofacial Surgery 23: 17-23.

 

Ennis, W. et al. (2005). "Ultrasound therapy for recalcitrant diabetic foot ulcers: results of a randomized, double-blind, controlled, multicenter study." OstomyWound Management 51(8): 24-26.

 

Enwemeka, C. S. (1989). "The effects of therapeutic ultrasound on tendon healing. A biomechanical study." Am J Phys Med Rehabil 68(6): 283-7.

 

Enwemeka, C. S. et al. (1990). "The biomechanical effects of low-intensity ultrasound on healing tendons." Ultrasound Med Biol 16(8): 801-7.

 

Frizzell, L. A. and F. Dunn (1982). Biophysics of ultrasound. Therapeutic Heat and Cold. J. Lehmann. Baltimore, Williams & Wilkins.

 

Fyfe, M. C. and L. A. Chahl (1982). "Mast cell degranulation: A possible mechanism of action of therapeutic ultrasound (Abstract)." Ultrasound in Med & Biol 8(Suppl 1): 62.

 

Griffin, J. E. (1980). "Transmissiveness of ultrasound through tap water, glycerin, and mineral oil." Phys Ther 60(8): 1010-1016.

 

Gomez, M. A. et al. (1989). "Medical collateral ligament healing subsequent to different treatment regimens." J. Appl. Physiol. 66(1): 245-252.

 

Gutiérrez, M. I. et al. (2010). Methods for Characterization of Physiotherapy Ultrasonic Transducers. New Developments in Biomedical Engineering. D. Campolo, In Tech: 518.

 

Harris, F. et al. (2014). "Sounding the death knell for microbes?" Trends Mol Med 20(7): 363-367.

 

Harvey, W. et al. (1975). "The stimulation of protein synthesis in human fibroblasts by therapeutic ultrasound." Rheumatol Rehabil 14(4): 237.

 

Hashish, I. et al. (1986). "Anti-inflammatory effects of ultrasound therapy: evidence for a major placebo effect." Br J Rheumatol 25(1): 77-81.

 

Hauser, R. et al. (2013). "Ligament Injury and Healing: A Review of Current Clinical Diagnostics and Therapeutics." Open Rehabilitation Journal 6: 1-20.

 

Hill, M. et al. (2003). "Muscle satellite (stem) cell actiation during local tissue injury and repair." J Anat 203: 89-99.

 

Hoogland, R. (1995). Ultrasound Therapy. Delft, Enraf Nonius.

 

Huys, S. et al. (1993). "Comparison of effects of early and late ultrasound treatment on tendon healing in the chicken limb." J Hand Ther 6: 58-59.

 

Johns, L. D. et al. (2007). "Analysis of effective radiating area, power, intensity, and field characteristics of ultrasound transducers." Arch Phys Med Rehabil 88(1): 124-9.

 

Khanna, A., R. T. Nelmes, et al. (2009). "The effects of LIPUS on soft-tissue healing: a review of literature." Br Med Bull 89: 169-182.

 

Khatri, S. et al (2008). "Effectiveness of therapeutic ultrasound in identification of tibial stress fractures." Indian Journal of Physiotherapy and Occupational Therapy-An International Journal 2(1): 51-53.

 

Kitchen, S. and C. Partridge (1997). "Ultrasound, shortwave diathermy and laser: a survey to examine patterns of use in England." Br J Therapy Rehabil 4(2): 75-78.

 

Kitchen, S. and C. J. Partridge (1990). "A review of therapeutic ultrasound:

Part 1 - Background and physiological effects

Part 2 - The efficacy of ultrasound

Part 3 - References." Physiotherapy 76(10): 593-600.

 

Klucinec, B. et al. (2000). "Effectiveness of wound care products in the transmission of acoustic energy." Phys Ther 80(5): 469-76.

 

Latey, P. (2001). "Aspects of inflammation." J Bodywork and Movement Therapies 5(2): 124-131.

 

Lehmann, J. (1982). Therapeutic Heat and Cold. Baltimore, Williams & Wilkins.

 

Lehmann, J. F. and B. J. de Lateur (1989). Ultrasound, shortwave, microwave, superficial heat & cold in the treatment of pain. Textbook of Pain. P. D. Wall and R. Melzack. Edinburgh, Churchill Livingstone: 932-941.

 

Leonard, J. et al. (2004). "A comparison of intramuscular temperatures during 10-minute 1.0-MHz ultrasound treatments at different intensities." J. Sport Rehabil. 13(3): 244-254.

 

Leung, M. C. et al. (2004). "Effect of ultrasound on acute inflammation of transected medial collateral ligaments." Arch-Phys-Med-Rehabil. 85(6): 963-6.

 

Leung, M. C. et al. (2006). "Therapeutic ultrasound enhances medial collateral ligament repair in rats." Ultrasound Med Biol 32(3): 449-52.

 

Levine, D. and T. Watson (2014). Therapeutic Ultrasound. Chapter 19 in : Canine Rehabilitation and Physical Therapy. D. L. Mills and D. Levine. Philadelphia, Elsevier/Saunders. ISBN : 978-1-4377-0309-2

 

Li, J. et al. (2003). "Cytokine release from osteoblasts in response to ultrasound stimulation." Biomaterials 24(13): 2379-2385.

 

Li, J. et al. (2002). "Optimum intensities of ultrasound for PGE(2) secretion and growth of osteoblasts." Ultrasound Med. Biol. 28(5): 683-690.

 

Lorena, D. et al. (2002). "Nornal scarring : importance of myofibroblasts." Wound Repair Regen 10(2): 86-92.

 

Lovric, V. et al. (2013). "The effects of Low-intensity Pulsed Ultrasound on tendon-bone healing in a transosseous-equivalent sheep rotator cuff model." Knee Surg Sports Traumatol Arthrosc 21(2): 466-475.

 

Majlesi, J. and H. Unalan (2004). "High-power pain threshold ultrasound technique in the treatment of active myofascial trigger points: a randomized, double-blind, case-control study." Arch Phys Med Rehabil 85(5): 833-836.

 

Manca, A. et al. (2014). "Ultrasound and laser as stand-alone therapies for myofascial trigger points: a randomized, double-blind, placebo-controlled study." Physiother Res Int 19(3): 166-175.

 

Markert, C. D. (2004). "Ultrasound and exercise in skeletal muscle regeneration."

 

Maxwell, L. (1992). "Therapeutic ultrasound: Its effects on the cellular & mollecular mechanisms of inflammation and repair." Physiotherapy 78(6): 421-426.

 

McBrier, N. M. et al. (2007). "Therapeutic ultrasound decreases mechano-growth factor messenger ribonucleic acid expression after muscle contusion injury." Arch Phys Med Rehabil 88(7): 936-40.

 

McDiarmid, T. et al. (1985). "Ultrasound and the treatment of pressure sores." Physiotherapy 71(2): 66-70.

 

Meakins, A. and T. Watson (2006). "Longwave ultrasound and conductive heating increase functional ankle mobility in asymptomatic subjects." Physical Therapy in Sport 7: 74-80.

 

Morishita, K. et al. (2014). "Effects of therapeutic ultrasound on range of motion and stretch pain." J Phys Ther Sci 26(5): 711-715.

 

Mortimer, A. J. and M. Dyson (1988). "The effect of therapeutic ultrasound on calcium uptake in fibroblasts." Ultrasound in Med & Biol 14(6): 499-506.

 

Naghdi, S. et al. (2008). "A Clinical Trial on the Treatment of Chronic Rhinosinusitis with Continuous Ultrasound." Journal of Physical Therapy Science 20(4): 233-238.

 

Neidlinger-Wilke, C. et al. (2002). "Fibroblast orientation to stretch begins within three hours." J Orthop Res 20(5): 953-6.

 

Ng, G. Y. F. (2011). "Comparing therapeutic ultrasound with microamperage stimulation therapy for improving the strength of Achilles tendon repair." Connective Tissue Research 52(3): 178-182.

 

Nussbaum, E. (1998). "The influence of ultrasound on healing tissues." Journal of Hand Therapy 11(2): 140-7.

 

Nussbaum, E. L. (1997). "Ultrasound: to heat or not to heat - that is the question." Physical Therapy Reviews 2: 59-72.

 

Nussbaum, E. L. and M. Locke (2007). "Heat shock protein expression in rat skeletal muscle after repeated applications of pulsed and continuous ultrasound." Arch Phys Med Rehabil 88(6): 785-90.

 

Paliwal, S. and S. Mitragotri (2006). "Ultrasound-induced cavitation: applications in drug and gene delivery." Expert Opin Drug Deliv 3(6): 713-726.

 

Papalada, A. et al (2012). "Ultrasound as a primary evaluation tool of bone stress injuries in elite track and field athletes." Am J Sports Med 40(4): 915-919.

 

Polak, A. et al. (2014). "A prospective, randomized, controlled, clinical study to evaluate the efficacy of high-frequency ultrasound in the treatment of Stage II and Stage III pressure ulcers in geriatric patients." Ostomy Wound Manage 60(8): 16-28.

 

Poltawski, L. and T. Watson (2007). "Relative transmissivity of ultrasound coupling agents commonly used by therapists in the UK." Ultrasound Med Biol 33(1): 120-8.

 

Pope, G. D. et al. (1995). "A survey of electrotherapeutic modalities: ownership and use in the NHS in England." Physiotherapy 81(2): 82-91.

 

Pye, S. (1996). "Ultrasound therapy equipment - does it perform?" Physiotherapy 82(1): 39-44.

 

Ramirez, A. et al. (1997). "The effect of ultrasound on collagen synthesis and fibroblast proliferation in vitro." Medicine and Science in Sports and Exercise 29(3): 326-32.

 

Reher, P. et al. (1999). "Effect of ultrasound on the production of IL-8, basic FGF and VEGF." Cytokine 11(6): 416-23.

 

Robertson, V. J. et al. (2006). Electrotherapy Explained: Principles and Practice. Oxford, Elsevier.

 

Sahu, N. et al (2019) Continuous low-intensity ultrasound attenuates IL-6 and TNFalpha-induced catabolic effects and repairs chondral fissures in bovine osteochondral explants BMC

Musculoskelet Disord 20(1): 193

 

Sarrafzadeh, J. et al. (2012). "The effects of pressure release, phonophoresis of hydrocortisone, and ultrasound on upper trapezius latent myofascial trigger point." Arch Phys Med Rehabil 93(1): 72-77.

 

Schabrun, S. et al. (2006). "Are therapeutic ultrasound units a potential vector for nosocomial infection?" Physiother Res Int 11(2): 61-71.

 

Serena, T. et al. (2009). "The impact of noncontact, nonthermal, low-frequency ultrasound on bacterial counts in experimental and chronic wounds." Ostomy Wound Manage 55(1): 22-30

 

Shah, S. G. S. and A. Farrow (2012). "Trends in the availability and usage of electrophysical agents in physiotherapy practices from 1990 to 2010: a review." Physical Therapy Reviews 17(4): 207-226.

 

Spratt, H. et al. (2014). "Physical therapy clinic therapeutic ultrasound equipment as a source for bacterial contamination." Physiother Theory Pract 30(7): 507-511.

 

Srbely, J. Z. and J. P. Dickey (2007). "Randomized controlled study of the antinociceptive effect of ultrasound on trigger point sensitivity: novel applications in myofascial therapy?" Clin Rehabil 21(5): 411-7.

 

Stanisic, M. et al. (2005). "Wound debridement with 25 kHz ultrasound." Adv Skin Wound Care 18(9): 484-490.

 

Straub, S. J. et al. (2008). "Variability in effective radiating area at 1 MHz affects ultrasound treatment intensity." Phys Ther 88(1): 50-7; discussion 58-62.

 

Sutton, A, and Watson, T. (2011) Electrophysical Agents in Physiotherapy Ch 93 IN : Diagnosis and Management of Lameness in the Horse Ed : MW Ross and SJ Dyson Pub : Elsevier / Saunders (2nd Edition)

 

ter Haar, G. (1987). Recent advances in therapeutic ultrasound. Ultrasound: Method Applications, Biological Effects and Hazzard Potential. M. Repacholi, M. Grandolfo and A. Rindi. New York, Plenum Press.

 

ter Haar, G. (1999). "Therapeutic Ultrsound." Eur J Ultrasound 9: 3-9.

 

Toumi, H. and H. M. Best (2003). "The inflammatory response: friend or enemy for muscle injury?" Br J Sports Med 37: 284-286.

 

Tsai, W. C. et al. (2006). "Ultrasound stimulation of types I and III collagen expression of tendon cell and upregulation of transforming growth factor beta." J Orthop Res 24(6): 1310-6.

 

Tsai, W. C. et al. (2011). "Effect of therapeutic ultrasound on tendons." Am J Phys Med Rehabil 90(12): 1068-1073.

 

Turner, S. et al. (1989). "The effect of ultrasound on the healing of repaired cockrel tendon: is collagen cross-linkage a factor?" J Hand Surg 14B: 428-433.

 

Uhlemann, C. et al. (2003). "Therapeutic ultrasound in lower extremity wound management." International Journal of Lower Extremity Wounds 2(3): 152-157.

 

Unalan, H. et al. (2011). "Comparison of high-power pain threshold ultrasound therapy with local injection in the treatment of active myofascial trigger points of the upper trapezius muscle." Arch Phys Med Rehabil 92(4): 657-662.

 

Velnar, T. et al. (2009). "The Wound Healing Process: an Overview of the Cellular and Molecular Mechanisms." The Journal of International Medical Research 37(5): 1528-1542.

 

Wang, E. D. (1998). "Tendon repair." J Hand Ther 11(2): 105-10.

 

Ward, A. R. (1986). Electricity, Fields and Waves in Therapy. Marrickville, Australia, Science Press.

 

Warden, S. J. et al. (2006). "Low-Intensity Pulsed Ultrasound Accelerates and a Nonsteroidal Anti-inflammatory Drug Delays Knee Ligament Healing." Am J Sports Med 34(7): 1094-102.

 

Watson, T. (2000). "Masterclass. The role of electrotherapy in contemporary physiotherapy practice." Manual Therapy 5(3): 132-41.

 

Watson, T. (2002). "Ultrasound Dose Calculations." In Touch 101: 14-17.

 

Watson, T. (2003). "Soft Tissue Healing." In Touch 104: 2-9.

 

Watson, T. (2006). "Electrotherapy and tissue repair." Sportex-Medicine. 29: 7-13.

 

Watson, T. (2006). "Tissue repair: The current state of the art." Sportex-Medicine. 28: 8-12.

 

Watson, T. (2008). "Ultrasound in contemporary physiotherapy practice." Ultrasonics. 48(4): 321-329.

 

Watson, T. and S. Young (2008). Therapeutic Ultrasound. Electrotherapy : Evidence Based Practice. T. Watson. Edinburgh, Churchill Livingstone - Elsevier.

 

Watson, T. (2010). Narrative Review : Key concepts with electrophysical agents Physical Therapy Reviews 15(4): 351-359.

 

Watson, T. (2014). "Crest of a Wave: Effectiveness of therapeutic ultrasound in musculoskeletal injury." International Therapist(110): 18-20.

 

Watson, T. (2020) Therapeutic Ultrasound Ch 9 IN : Electrophysical Agents: Evidence Based Practice. Elsevier, Oxford

 

Werner, S. and R. Grose (2003). "Regulation of wound healing by growth factors and cytokines." Physiol. Rev 83: 835-870.

 

Wilkin, L. D., et al. (2004). "Influence of therapeutic ultrasound on skeletal muscle regeneration following blunt contusion." Int. J. Sports Med. 25(1): 73-77.

 

Williams, R. (1987). "Production and transmission of ultrasound." Physiotherapy 73(3): 113-116.

 

Yavuz, F. et al. (2014). "Low-level laser therapy versus ultrasound therapy in the treatment of subacromial impingement syndrome: A randomized clinical trial." J Back Musculoskelet Rehabil 27(3): 315-320.

 

Yeung, C. K. et al. (2006). "Pulsed ultrasound treatment accelerates the repair of Achilles tendon rupture in rats." J Orthop Res 24(2): 193-201.

 

Young, S. R. and M. Dyson (1990). "The effect of therapeutic ultrasound on angiogenesis." Ultrasound Med Biol 16(3): 261-9.

 

Young, S. R. and M. Dyson (1990). "Macrophage responsiveness to therapeutic ultrasound." Ultrasound Med Biol 16(8): 809-16.

 

Zeng, C. et al. (2014). "Effectiveness of continuous and pulsed ultrasound for the management of knee osteoarthritis: a systematic review and network meta-analysis." Osteoarthritis Cartilage 22(8): 1090-1099.

 

Zhou, S. et al. (2004). "Molecular mechanisms of low intensity pulsed ultrasound in human skin fibroblasts." J Biol Chem 279(52): 54463-9.

 

Zura, R., S. Mehta, et al. (2015). "A cohort study of 4,190 patients treated with low-intensity pulsed ultrasound (LIPUS): findings in the elderly versus all patients." BMC Musculoskeletal Disorders 16(1): 45.

bottom of page